Sample Throughput: 1~8 samples at the same time
Pipetting Range: 20μL ~250μL
Detection Time: 1~2 hours,relying on the reagent
Channel and Available Fluorescein:
- Channel 1: FAM, SYBR Green I, etc.
- Channel 2: VIC, HEX, TET, JOE, etc.
- Channel 3: ROX, Texas Red, etc.
- Channel 4: Cy5, etc.
Dimension: 750mm(L) × 350mm (W) × 600mm (H)
Weight: 80Kg
Pipetting Performance:
- 20μL≤V<40μL: accuracy:A≤5.0%,repeatability:CV≤3.0%
- 40μL≤V<100μL: accuracy: A≤3.0%,repeatability: CV≤1.5%
- V≥100μL: accuracy:A≤1.0%,repeatability:CV≤1.0%
Extraction Heating Rate: Average heating rate: ≥1.5℃/s
Extraction Temperature Accuracy: ≤1.0℃
PCR Heating Rate:
- Average heating rate:≥ 4.5℃/s
- Maximum heating rate:≥ 6.1℃/s
PCR Cooling Rate:
- Average cooling rate: ≥ 3.5'℃/s
- Maximum cooling rate:≥ 5.0℃/s
PCR Temperature Accuracy: ≤0.1%
Sample Testing Linearity and Repeatabilit:
- Linear correlation: /r/≥ 0.998
- Repeatability: CV ≤1.5%
Information Management:
- Sample information: with scanner inside
- Panall 8000 can scan and record the sample information
- Reagent information: visual system can automatically identify
- the kit information and run the corresponding program
Minimized Contamination: Directional exhaust & negative pressure system, HEPA fltration,UV disinfection, Shortest fixed path for sample operation
Data Storage: 1000 experiment files can be stored
Language: Chinese and English
Power Supply and Power Consumption: AC 100~240V,50/60Hz:1200VA
Communication Specification: Internet Port: TCP/lP protocol, Ethernet connection, USB Port: USB 2.0