new products

ZY-400 Fully Automatic Clinical Chemistry Analyzer

ZY-400 Fully Automatic Clinical Chemistry Analyzer

Automated Washing System:

  • Permanent glass cuvettes with good light transmission, anti-corrosion, anti-wear, and durable
  • Multi-step warm water washing system with more stabletemperature of the cuvette tray and more thoroughly washedcuvettes
  • Independently operating, highly polished, three-dimensional vortex-style automatic washing system
  • Equipped with KHB special packaging detergent, with which maintenance costs are reduced andcross-contamination rate is lower than 0.08%

Smart Filling System (Sample/Reagent Probes):

  • Liquid-level detection and liquid volume tracking function
  • Anti-collision function (optional) and sample clot detection
  • Agglomeration monitoring alarm function (optional)

Excellent Overall Analyzer Performance:

  • Able to function for 24 hours; automatic "sleep" and "wake-up" functions
  • General samples can be reclassified as emergency samples for priority detection
  • 13 optional pre-set report formats, or user-defined report formats
  • Duplex communication between the device and external computer systems (LIS/HIS) through data interface and Internet protocol is supported
  • The user analysis function is provided for data security

Performance:

Througput: Photometric test: 400 tests/hour, ISE: 240 tests/hour (Na+, K+, Cl-, Ca2+)

System functions: Continuous standby for 24 hours, emergency test, automatic dilution, automatic retest, automatic sleep and wake-up Single/dual reagent (s) and wavelength (s) settings, 1-10 min response time

Type of sample: For whole blood test, serum test, plasma test, urine test, among others, please see Instructions for Use

Analytical method: One-point endpoint method, dual-point endpoint method, reaction speed method (kinetic method), two-point speed method (fixed time method) and potential survey method for ISE test

Calibration method: Linear calibration: factor K method, linear method (including two-point linear and multi-point linear methods) Non-linear calibration: spline method, broken line method, exponential method, logarithmic method, 3P parameter method, 4P parameter method and 5P parameter method

Sample System:

Sample tray: 90 sample positions available for mixed placement of miscellaneous samples and bar code scanning (optional)

Sample volume: 1.5 ~ 70µL, progressively increase by 0.1µL

Sample probe: Liquid level detection, volume tracking, automatic anti-collision protection, probe blockage detection (optional), etc. Inner and outer wall cleaning technology, automatic maintenance and cleaning, with a cross-contamination rate of lower than 0.08%

Sample tube: A variety of specifications for micro sample cup and sample tubes available on market

Reagent System:

Reagent tray: 72 reagent positions, 36 positions on both inner and outer rings respectively, available for 72 reagent bottles with volume of 35ml and 65ml. Scalable bottle volume, allow bar code scanning (optional) Cleaning solution settable at any position

Reagent volume: 10 ~ 350µL, progressively increase by 0.5µL

Reagent probe: Liquid level detection, volume tracking, automatic anti-collision protection Inner and outer wall cleaning technology, automatic maintenance and cleaning, with a cross-contamination rate of lower than 0.08%

Temperature for refrigeration of reagent: Continuous refrigeration for 24 hours at 2 ~ 8℃

Reaction System:

Reaction tray: Tray design, with 81 cuvettes

Cuvette: Permanent hard glass cuvette (glass) with optical path of 6mm, semi-permanent organic glass (plastic) (optional)

Reaction solution volume: 110 ~ 500µL

Constant temperature system: Liquid-gas synchronous constant temperature device

Temperature of reaction tray: 37℃, real-time display of temperature

Temperature fluctuation of reaction tray: ±0.1℃

Stirring mechanism: Independent stirrer

Optical System:

 Light source: Halogen lamp, available for automatic sleep and wake-up

Spectrophotometry: Concave grating photometry, post-spectroscopic technology

Wavelength: 12 wavelengths: 340, 380, 405, 450, 505, 546, 578, 600, 660, 700, 750, 800

Wavelength precision: ≤±1nm

Absorbance range: -0.5 ~ 3.5Abs

Working Conditions:

Power supply: 230V±23, 50/60 Hz

Power: < 1500VA (UPS 2000VA power supply is optional)

Water consumption: Less than 10 L/H deionized water

Dimensions: 1075mm (length) ×706mm (width) ×1100mm (height)