Performance:
Throughput: Biochemical constant test speed: 200 tests/hour
System functions: Continuous standby for 24 hours, emergency test, automatic dilution, automatic retest, automatic sleep and wakeup Single/dual reagent (s) and wavelength (s) settings, 1 ~ 15 min response time
Type of sample: For whole blood test, serum test, plasma test, urine test, among others, please see Instructions for Use
Analytical method: One-point endpoint method, dual-point endpoint method, reaction speed method (kinetic method), two-point speed method (fixed time method) and potential survey method for ISE test
Calibration method: method (fixed time method) and potential survey method for ISE test Linear calibration: factor K method, linear method (including two-point linear and multi-point linear methods) Non-linear calibration: spline method, broken line method, exponential method, logarithmic method, 3P parameter method, 4P parameter method and 5P parameter method
Sample System:
Sample tray: 54 sample positions available for mixed placement of miscellaneous samples and bar code scanning (optional)
Sample volume: 2.5 ~ 70µL, progressively increase by 0.1µL
Sample probe: Liquid level detection, volume tracking, automatic anti-collision protection, probe blockage detection (optional), etc. Inner and outer wall cleaning technology, automatic maintenance and cleaning, with a cross-contamination rate of lower than 0.08%
Sample tube: A variety of specifications for micro sample cup and sample tubes available on market
Reagent System:
Reagent tray: 72 reagent positions, 36 positions on both inner and outer rings respectively, available for 72 reagent bottles with volume of 20mL and 35mL. Scalable bottle volume, allow bar code scanning (optional) Cleaning solution settable at any position
Reagent volume: 20 ~ 350µL, progressively increase by 0.5µL
Reagent probe: Liquid level detection, volume tracking, automatic anti-collision protection Inner and outer wall cleaning technology, automatic maintenance and cleaning, with a cross-contamination rate of lower than 0.08%
Temperature for refrigeration of reagent: Continuous refrigeration for 24 hours at 2 ~ 8℃
Reaction System:
Reaction tray: Tray design, with 80 cuvettes
Cuvette: Semi-permanent organic glass (plastic) with optical path of 6mm
Reaction solution volume: 120 ~ 500µL
Constant temperature system: electrical heating constant temperature system
Temperature of reaction tray: 37℃, real-time display of temperature
Temperature fluctuation of reaction tray: ±0.1℃
Stirring mechanism: Independent stirrer
Optical System:
Light source: Halogen lamp, available for automatic sleep and wakeup
Spectrophotometry: Concave grating photometry, post- spectroscopic technology
Wavelength: 12 wavelengths: 340,380,405,450,505,546,578,600,660,700,750,800
Wavelength precision: <±1nm
Absorbance range: -0.5 ~ 3.3 Abs
Working Conditions:
Power supply: 230V±23, 50/60Hz
Power: 900VA (UPS 1500VA power supply is optional)
Water consumption: Less than 7 L/H deionized water
Dimensions: 890mm (length) X725mm (width) X 550mm (height)
Weight: 100kg