Performance:
Througput: Photometric test: 800 tests/hour
System functions: Consecutive 24-hour operation. STAT sample priority, auto pre-dilution, serum index, remote diagnosis, auto
Type of sample: Whole blood, serum, plasma, urine, cerebrospinal fluid, hydrothorax and ascites etc.
Analytical method: One Point Method, Two Point Method, Kinetic Method, Fixed-time Method, ISE Potential Direct Measurement Method
Expanded Function: Online reagent loading function; enzyme linear expansion function; quantitative analysis of serum indices (lipidemia, hemolysis, jaundice)
Software Connection: Connectable to HIS, LIS systems
Sample System:
Sample Capacity: 180 positions for various samples, extendable, dual circular disc design, and support barcode scanning (optional). 60 positions are calibration/QC with refrigeration (optional), and different samples can be mixed in positions
Sample volume: 1 ~ 70µL, step by 0.1µL increasing
Sample probe: Liquid level detection, residual monitoring, collision protection, clot detection (optional), etc.
Sample tube: Micro sample cup, blood collection tube, plastic tube, etc.
Reagent System:
Reagent Capacity: 72 reagent positions, 36 positions on both inner and outer rings respectively, available for 72 reagent bottles with volume of 35ml and 65ml. Scalable bottle volume, allow bar code scanning (optional) Cleaning solution settable at any position
Reagent volume: 10 ~ 350µL, progressively increase by 0.5µL
Reagent probe: Liquid level detection, residual monitoring, collision protection, clot detection, etc.
Reagents Refrigeration: Allow consecutive 24-hour refrigeration at 2 ~ 8℃ temperature
Reaction System:
Reaction disc: Rotating disk, 160 cups
Cuvette: Semi-permanent plexiglass cuvette(plastic), optional permanent hard glass cuvette (glass); optical length: 5mm
Reaction Volume: 90 ~ 550µL
Heating System: Automatic cycling liquid bath
Incubator Temperature: 37℃, real-time display
Temperature Fluctuation: ≤±0.1℃
Mixing System: Independent mix arm
Optical System:
Light source: Halogen lamp, , automatic sleep and wake-up
photometer: Concave grating photometry, post-spectroscopic technology
Wavelength: 16 wavelengths, 340 ~ 800nm
Wavelength precision: ≤±1nm
Absorbance measurement range: -0.5 ~ 3.5Abs
Operating Conditions:
Power supply: 230V±23V, 50/60Hz
Input Supply: < 1500VA (UPS 2000VA power supply is optional)
Water consumption: Less than 30 L/hr
Dimensions: 1508mm×865mm×1204mm
Weight: 400kg