Real-time PCR of Tianlong
Gentier 96 E/R

Model

Gentier 96R

Gentier 96E

Throughput

1-96

 

1-96

 

Fluorescence Channels

4

6

Fluorescence Scanning Time

7s

 

7s

 

Dye Compatibility

·        Channel 1: FAM, SYBR Green I , SYTO 9, Eva Green, LC Green

·        Channel 2: HEX, VIC, TET, JOE

·        Channel 3: ROX, Texas Red

·        Channel 4: Cy5

·        Channel 1: FAM, SYBR Green I , SYTO 9, Eva Green, LC Green

·        Channel 2: HEX, VIC, TET, JOE

·        Channel 3: ROX, Texas Red

·        Channel 4: Cy5

·        Channel 5: Alexa Fluor 680

·        Channel 6: FRET

Optical System

Light Source

High-brightness, long life and maintenance-free LED light source, excitation from the top

High-brightness, long life and maintenance-free LED light source, excitation from the top

Detector

Photodiodes (PDs), top scanning

 

Photodiodes (PDs), top scanning

Excitation Range

CH1: 465nm CH2:527nm CH3:580nm CH4:632nm

CH1: 465nm CH2:527nm CH3:580nm CH4:632nm CH5:680nm CH6:465nm

Detection Range

CH1: 510nm CH2:563nm CH3:616nm CH4:664nm

CH1: 510nm CH2:563nm CH3:616nm CH4:664nm CH5:730nm CH6:616nm

Fluorescence Dynamic Range

Adjustable

Adjustable

Sample Dynamic Range

From 101 to 1010copies

 

From 101 to 1010copies

 

Thermal Black

Heating Method

Peltier

Peltier

Heating Rate

≥6.1 ℃/s ℃

≥6.1 ℃/s ℃

Cooling Rate

≥5.2 ℃/s

 

≥5.2 ℃/s

 

Temperature Accuracy

≤0.1 ℃

≤0.1 ℃

Gradient Range

1℃-40 ℃

 

1℃-40 ℃

 

Gradient Block

12 row

12 row

Special Temperature Protocol

Thermal gradients PCR, long PCR, Touch Down PCR

Thermal gradients PCR, long PCR, Touch Down PCR

 

Sample Testing Linearity and Repeatability

Linear correlation: /r/ ≥ 0.999
Repeatability: cycle threshold (Ct) value CV≤0 0.5%

Linear correlation: /r/ ≥ 0.999
Repeatability: cycle threshold (Ct) value CV≤0 0.5%

Software Functions

Control Modes

Model 1: 10.4 inch touch screen
Mode 2: PC direct control

Model 1: 10.4 inch touch screen
Mode 2: PC direct control

Power Failure Protection

Automatically start running experiments after power supply, no need to wait PC software

Automatically start running experiments after power supply, no need to wait PC software

Data Storage and Transmission

Upload and download through USB disk. 1000 results can be stored in machine

Upload and download through USB disk. 1000 results can be stored in machine

Reporting Function

Templates reserved; customized experiment report

Templates reserved; customized experiment report

Key Applications

Relative quantification, absolute quantification, melting curve analysis, SNP analysis

Relative quantification, absolute quantification, melting curve analysis, SNP analysis

Others

Operating System for PC

Win7/Win10/Win11

 

Win7/Win10/Win11

 

Power Supply and Power Consumption

AC 100-240V,50-60HZ, 900VA

AC 100-240V, S0-60HZ; 600VA

Weight

30 Kg (net)

 

30 Kg (net)

 

Instrument Dimension

555*475*484mm (W*L*H)

 

555*475*484mm (W*L*H)

Suitable Consumables

0.2mL 96-well plates, 8-tube strips, single tubes(clear, frosted and white).

0.2mL clear non-skirt 48-WELL plates, 8>tube strips, single tubes, etc.

 

Certificate

IVDR CE